logo
Hamilton logo
How a dispenser works
spacer
Dispenses, including titrations, can be made in two easy steps (after priming the instrument) as shown in Figure 1.
 
Figure 1.
Figure 1: A simple, dual dispense
Step 1: Fill the syringe(s) with the programmed amount of reagent from the reservoir.

Step 2: Dispense the programmed amount(s) into a microwell plate, test tube, vial, etc. to complete the dispense cycle.
pixel
In addition to simple dispenses performed in all MICROLAB 500 dispensers, models ML510B, ML540B, ML511C, ML541C and ML560C can also perform aliquot, serial, pipette and titrate dispenses. See Figure 2.
 
ML 500 Series Method Illustration
Method Type A B C
Simple dispense disp1 Reagent is filled from a reservoir and dispensed
Aliquot dispense   disp2 Repetitive dispense with constant final volume
Serial dispense   disp3 Repetitive dispense with a variable final volume
Pipette   disp4 Aspirate sample, then dispense
Titrate   disp5 Repetitive dispense to endpoint
Reagent dilution
(dual syringe only)
disp6 Addition of two reagents to a common vial
Return to reservoir   disp7 Save reagent in the fluid path by returning to reagent reservoir
Figure 2:
Dispense method types
 Blue = diluent
 Yellow = sample
 Green = diluted sample
 Purple = titration endpoint