LTN
Luer Tip Cemented Needle |
Needles
are cemented into the conical glass luer tip
of the barrel at a point corresponding to the
zero graduation mark. |
| |
 |
N,NR
Cemented |
Needles are cemented
into the glass syringe barrel at a point corresponding
to the zero graduation mark. NR stands for syringes
with Rheodyne specified needles. |
| |

|
SN,
LTSN
Special Needle, Luer Tip Special Needle |
Hamilton offer
customised syringes with special needles. State
the following: needle length, guage, point style,
and whether electro-tapered. |
| |
 |
LT
Luer Tip |
Ground glass syringe barrel
with a male luer taper accepts most hyperdermic
needles. Use Kel-F hub needle and connectors
for a tight seal. |
| |
 |
RN,
RNR
Removable Needle, Removable Rheodyne Needle |
Removable needles
seat precisely to the zero mark of the syringe.
Allows the use of different specification needles
on the same syringe barrel. RNR stands for syringes
with Rheodyne specified removable needles. |
| |
 |
KH
Knurled Hub |
Replaceable needle;
knurled hub is used on 7000 series syringes
exclusively. The attachment of a spacer enables
repeatable depth injections. |
| |
 |
SL
SampleLock |
On/Off syringe
valve with RN needle is cemented to a syringe
barrel. Used for headspace, environmental sample
collection and storage, pre-pressurisation of
gaseous samples for GC analysis, and sample
spiking. |
| |
 |
TLL,
TLLX
Teflon® Luer Lock |
Male luer taper
with nickel-plated brass hub accepts and locks
in place luer hub needles and connectors. The
X-style plunger stop incorporates a 6-32 UNC
female thread on the end of the stop to allow
attachment to drive arm mechanisms such as on
the MICROLAB 500 series diluters/dispensers. |
| |

|
TLLXB
TLL with Stop & Backflushing |
Used in drive
arm mechanisms such as on MICROLAB 4000 or PSD2.
Back-bushing syringes increase plunger tip life,
maintain proper plunger alignment and reduce
the particles generated by plunger to glass
contact. |
| |

|
TLL-XL
60mm TLL Syringe |
These 60mm stroke
TLL-style syringes are compatible with many
liquid handling instruments, including XL-3000
modular digital syringe pump, Tecan, Rosys,
Cavro IQ series sample processors and OEM liquid
handling applications. |
| |
 |
TLL/TLLX
Short Stroke |
These syringes
are designed specifically for the PSD3 (half
height syringe pump). |
| |
 |
C/CX
Cheminert |
1/4"-28
UNF, male fitting. Used in low volume applications
where system dead volumes need to be minimised.
The X-style plunger stop incorporates a 6-32
UNC female thread on the end of the stop to
allow attachment to drive arm mechanisms, such
as on the MICROLAB 500 series diluters/dispensers. |
| |
 |
C-XP
30mm C Syringe |
These 30mm stroke
C-style syringes are compatible with many liquid
handling instruments, including XL-3000 modular
digital syringe pump, Tecan, Rosys, Cavro IQ
series sample processors and OEM liquid handling
applications. |
| |
 |
DX
Dilutor |
TLL, TEFLON Luer
Lock, male fitting with M6 x 1 female, side
port. The X-style plunger stop incorporates
a 6-32 UNC female thread on the end of the stop
to allow attachment to drive arm mechanism,
such as on the MICROLAB 500 series diluters/dispensers. |
| |
 |
AD
AccuDil® |
M8 x 0.75, male
fitting. Used with MICROLAB 1000 plus diluter/dispenser. |
| |
 |
DAD
Diluter AccuDil® |
M8 x 0.75 male
fitting with M6 x 1 sideport. Used with MICROLAB
1000 plus diluter/dispenser. |
| |
 |
Electro-tapered |
Some replacement
needles can be specified as electro-tapered.
Use this style for applications such as attaching
PTFE tips, or for use on on-column GC injectors. |