Reliable Measurements 10. Contamination From the Analyst and Apparatus By Paul Gaines, Ph.D. • Edited by Brian Brolin
Part 8, Environmental Contamination,
discussed the importance of a clean room to reduce air contaminants
while performing trace analyses. A clean room is equally important for
minimizing contamination from the analyst.
The
area pictured in figure 10.1 is the buffer zone between the outside air
and the main clean room. Air entering with the analyst is filtered (at
the rate of 1.5 times per minute) before the analyst enters the main
clean area. A cart contains a HEPA vacuum used for cleaning objects to
be taken into the main clean room. A ‘sticky mat' on the floor just
inside the room will remove shoe dust and other contaminants. Another
sticky mat (not pictured) is placed just before the entry way to the
buffer zone.
Fig. 10.1: Entering the Clean Room
The
analyst pictured here is wearing disposable "clean room" clothing.
There are three separate pieces (head covering, foot covering, and main
coverall). The clothing is worn over normal street clothes and should
never leave the clean room area.
Common Contaminants:
Sweat contains Ca, Mg, Pb, K, NH4+ , SO4-2, PO4-3, and Cd (for those who smoke).
Cosmetics can contain high concentrations of Al, Be, Ca, Cu, Cr, K, Fe, Mn, Ti, and Zn.
Some hair dyes contain Pb(OAC)2.
Dandruff shampoo can contain significant levels of Se.
Eye make-up may contain Hg as a preservative.
Calamine lotion is almost pure ZnO.
Watches and jewelry contain an assortment of elements and should not be worn in the laboratory.
Tips for the Analyst:
Keep the volume of reagents to a minimum.
Keep
sample preparation and measurement times to a minimum (use timers and
don't let a digestate sit overnight or even during a lunch break).
Keep a history of beakers and other apparatus as to chemical exposure.
Exercise
care in handling reagents (exposure to atmosphere or pouring liquid
back into the same container or into the wrong container can result in
serious contamination).
Disposable powder-free
gloves prevent contamination from earlier exposures as opposed to
reusable gloves (vinyl is better for difficult manipulations such as
pipeting and handling wet glassware while latex is more rugged).
Disposable laboratory coats prevent contamination from earlier exposures.
Disposable foot coverings keep any dirt that the ‘sticky mat' missed contained.
NOTE: Dangerous operations require appropriate laboratory clothing REGARDLESS of the potential for contamination.
Another
potential source of contamination is casued by the apparatus that comes
into contact with the sample. Careful planning is required by the
analyst as to the type of apparatus that comes into contact with a
particular trace metals analysis. All apparatus should be specified in
the procedure along with any cleaning procedures or special
precautions. Although the type and degree of contamination that occurs
when using certain apparatus can be predicted, the importance of one or
more blanks with every preparation should be apparent from the
following table.
Table 10.1: Results of Spectrochemical Analysis (Hydrogen
Fluoride, Hydrogen Chloride, and Nitric Acid [all high purity] analyzed
after evaportation in Teflon, Platinum, and Quartz dishes)
Elements determined, ng/g ND = Not detected TR = Trace, not evaluated quantitatively
A Closer Look at Quartz:
Quartz
is a popular material for apparatuses used in trace elemental analysis.
There are two types of quartz: 1) opaque and 2) transparent.
Opaque quartz has the highest trace element concentration and should not be used for trace analysis. Transparent quartz
(types I and II) are made from naturally occurring quartz crystals or
sands. Type I is made by electric melting and type II by flame melting.
Type II has slightly less impurities than type I (some impurities are
volatilized by the flame). Type III quartz is made synthetically by
vapor phase hydrolysis of pure silicon compounds such as SiCl4. This type of quartz is more pure than the natural quartz with the exception of Cl - which is ~ 50 ppm. Type IV quartz is synthetically made from SiCl4
using a process involving electrical fusion of the oxidized staring
material. It is as pure as type III, with respect to trace metal
content, and contains much more Cl-. Use synthetic quartz whenever possible.
Table 10.2 shows typical impurities in natural and synthetic quartz.
Table 10.2: Recorded Impurity Levels in Types of Quartz1
ND = Not detected
Table 10.3 shows a comparison of impurities in natural quartz to other common laboratory apparatus.
Table 10.3: Trace Element Concentration of Container Materials
Figure
10.2 demonstrates how quartz is the cleanest of all sample container
materials that can withstand typical ashing temperatures (T > 400
�C).
Figure 10.2: Container Materials in order of Increasing Impurites1
Polyethylene - Low density (free radical initiated)
Fluorocarbons - (i.e. - Teflon, Tefzel, Halar, Kel-F)
Quartz - Synthetic
Polyethylene - High density
Quartz - Natural
Platinum
Borosilicate
Apparatus Tips:
Avoid
filtration due to contamination from the filter paper. It is far better
to centrifuge the sample using polyethylene tubes (acid-leached).
If
filtration is necessary, use Teflon filters as your first preference.
These are followed by cellulose acetate, polypropylene, or
polycarbonate filters that have been acid-leached.
Avoid rubber stoppers and all tubing except that made from Teflon or polyethylene.
For
grinding, use apparatus made from quartz, tungsten carbide, or pure
molybdenum since they introduce significant contamination of only a
single element (the tungsten carbide does introduce some cobalt).
When sieving, use nylon sieves.
Ovens and muffle furnaces are obvious sources of contamination and should be replaced with flames if at all possible.
Microwave
digestion in a closed Teflon (PFA) digestion vessel offers many obvious
advantages over other forms of sample preparation.
1. T. Murphy, National Bureau of Standards Special Publication 422, "Accuracy in Trace Analysis: Sampling, Sample Handling, and Analysis", Proceedings of the 7th IMR Symposium.